MEWOI1000-AC 0.0mA-1200A φ48mm H/L 60KV Voltage Clamp Meter

OEM/ODM Instruments Manufacturer

  • Main Features:: 1,current size: 48 mm,2,AC Current:0.0mA~1200A 3,AC Voltage:0.00 V~60KV (automatic shift),4,Data Storage: 99sets
  • Model No.:: MEWOI1000
  • Name:: MEWOI1000-AC 0.0mA-1200A φ48mm H/L 60KV Voltage Clamp Meter

Contact Us for QuotationEmail:mewoi@mewoi.comWhatsapp: +8618620373879

INTRODUCTION

Click to download:  MEWOI1000 Color Page

Click to download:  MEWOI1000 Manual

MEWOI1000.800.jpg


MEWOI1000 H/L Voltage Clamp Meter has unique features of automatic plug function, specially designed and manufactured for high voltage current measurement; use the latest CT technology and integrated mask digital technology; composed of special high voltage detector, wireless receiver and high voltage insulation rods. Wireless transmission can receive the measured data by penetrating storey building barriers or 300 meters. If not use insulated rods, it can also used as high-precision low-voltage clamp meter and leakage current meter, can accurately measure the 0.0mA ~1200A current or leakage current.  

The integrated design of clamp and lead zone of MEWOI1000 H/L Voltage Clamp Meter ensures uninterrupted year-round high precision, high reliability and high stability test. 

The connection rods of MEWOI1000 wireless high and low voltage clamp ammeter can be applied in high voltage line current measurement and online current measure below 60KV, also with the functions of peak value maintaining, data maintaining, data memory, etc. The special clamp ammeter can facilitate clamp to exit the tested lead by pressing or pulling back insulation bar, which is time efficient and convenient; widely used in power plant, substation, industrial and mining enterprise, inspection station, electrician maintenance department for current sensing and field working. It can also be used as high and low voltage changing ratio tester, that is, respectively test the first loop and secondary loop high and low voltage current, and then obtain the voltage change through manually calculation. Insulation rod has the characteristics of moisture resistance, high temperature resistance, impact resistance, bending resistance, high insulation and scalability, etc. 

MEWOI H/L Voltage Current Clamp Meter Selection Guide(Click the "Blue"Marked model No.to view the Spec)

Model No.Product NameAC RangeDefinitionClamp SizeData StorgeCircuit VoltageOther

MEWOI1000

MEWOI1000.Fluke.80.jpg

H/L Voltage Clamp Meter

AC  0.0mA~

1200A

0.0mA Automatic 

Shift 

φ50mm99setsbelow 60KV(operated with insulation rod,Bare wire35KV)

MEWOI1000C

MEWOI1000C.Extech.80.jpg

Wireless H/L Voltage Clamp Meterwireless(Distance:30M)

MEWOI1000T

MEWOI1000T.Extech.80.jpg

Portable H/L Voltage Clamp Meter

MEWOI1100

MEWOI1100.HIOKI.80.jpg

H/L Voltage Clamp Current MeterAC  0.00mA~600A0.01mAφ33mm99setsbelow 60KV(operated with insulation rod,Bare wire35KV)

MEWOI1100C

MEWOI1100C.Extech.80.jpg

Wireless H/L Voltage Clamp Current MeterAC  0.0mA~1200A0.1mAφ33mm99setsbelow 60KV(operated with insulation rod,Bare wire35KV)wireless(Distance:30M)
MEWOI1100T

MEWOI1100T.HIOKI.80.jpg

Portable H/L Voltage Clamp Current MeterAC  0.00mA~600A0.1mAφ33mm99setsbelow 10KV(operated with insulation rod,Bare wire 1KV)
MEWOI1200

MEWOI1200.Extech.80.jpg

H/L Voltage Clamp Current MeterAC  0.00mA~1000A(0.00mA~1600A Optional)0.01mAφ68mm99setsbelow 60KV(operated with insulation rod,Bare wire35KV)

MEWOI1200C

MEWOI1200C.80.jpg

Wireless H/L Voltage Clamp Current Meter99setswireless(Distance:30M)

TECHNICAL SPECIFICATION

Function

High/Low voltage AC current measurement, leakage current measurement, on-line AC current test, load current measurement

Power Supply

DC6V Alkaline Dry Battery (1.5V AAA X 4)

Size of jaw

Φ48mm

range

0.0mA~1200A

Resolution

0.1mA

Test Accuracy(23°C±5°C,Below 80%RH)

0mA~299mA: 1% 5dgt
0.30A~49.9A: 1.5% 5dgt
50.0A~199.9A: 2 5dgt
200A~600A: 3 5dgt
601A~1200A: 4% 5dgt

Frequency of current

50/60Hz Auto Identification

Insulation rod

300mm*3pcs

Test Method

Clamp CT, Integral Method

Display Mode

4 digital LCD display, backlight function, suitable for dark place.

LCD Dimension

47mm×28.5mm

Meter Dimension

76mmx255mmx31mm

Sampling Speed

About 2 times/second

Shift

Automatic Shift

Data Storage

99 Groups, during storage, the symbol “MEM” gives indication, “FULL” mark will flash to indicate the memory has been full.

Circuit Voltage

Test for the bare conductor below 35KV, test the current with insulation layer below 60kV(operated with insulation rod)

Peak Maintenance

Automatically maintain the height of the test value, usually in the test mode pressPEAK key, PEAK lights, that is, to keep open the PEAK function, then press PEAK key to cancel this function

Data Preservation

Under general test mode, press HOLD button to preserve data, “HOLD” symbol display. then press HOLD Key to cancel this function

Data Access

“MR” symbol indicates, it can read the stored data by turning upwards and downwards.

Overflow Display

Exceed measurement range overflow function: “OL A” symbol display

Auto Shut Down

About 15 minutes after booting, the meter will shut down automatically, in order to decrease the battery consumption.

Battery Voltage

When the battery voltage is below 4.8V, the symbol of battery voltage“    ”will show to remind to replace battery.

Weight

Tester: About 335g (Including battery and rod), total quality of meter: about 1.6Kg (including insulation rod and battery.)

Work Temperature and Humidity

-25°C ~40°C; Below 80%Rh

Storage Temperature and Humidity

-20°C ~60°C; Below 70%Rh

Insulation Intension

AC 3700V/rms. (Between insulation rod and clamp core)

Structure

Anti-Dripping Type II

Accessories

Meter:1pc; Meter Box:1pc; Insulation:1pc;Battery(Alkaline Dry Battery AAA):4pc.