| Contact Us for Quotation | Email:mewoi@mewoi.com | Whatsapp: +8618620373879 | 
| Products Guide Map: Click | 
MEWOI8042 Online Wireless Clamp on Current Monitoring Recorder
Ⅰ. INTRODUCTION
Click to download: MEWOI8042 User Manual
| MEWOI8042 Wireless Online Clamp Current Monitor Recorder use wireless communication to upload test data to concentrator or server for real-time monitoring of current, leakage current, troubleshooting of circuit insulation fault, monitoring of intermittent leakage fault, checking the time of insulation fault and fault location, etc. | 

MAIN FEATURES
| 1. The built-in communication module supports GPRS communication mode (nB-LOT /433/Lora communication mode can be customized), | 
| 2. Adopt the clamp structure  and no need to disconnect the tested circuit line, the test is safe and fast. Adopt double layer magnetic shield and digital integration technology, strong anti-interference and high precision. | 
| 3. With the function fo waterproof and dustproof, protection grade IP65. | 
| 4. With anti-theft function, equip with built-in electronic lock, control the clamp lock or unlock by infrared remote controller , can be installed in the outdoor and easy for using. | 
| 5. With LCDl display screen (display: current value, battery power, monitor number, network signal strength) can be used as a clamp ammeter, which is a multi-purpose machine. | 
| 6. Built-in large capacity lithium battery, ultra-long battery life, continuous online testing for 10 days.The monitor is equipped with a charging seat, which can charge 5 monitors once time. | 
Ⅱ. Technical Specification
| Monitor Model | MEWOI8042 | 
| Range | 0.0mA~300.0A | 
| Max.Measure Accuracy | ±1.5%±5dgt | 
| Max. Resolution | 0.1mA | 
| Clamp Size | Ф40mm | 
| Meter Size | 198x96x46mm | 
| Weight | 350g | 
| Communication Mode | GPRS communication mode (nB-LOT /433/Lora communication mode can be customized) | 
| Sampling Frequency | 1 times/second | 
| Upload Frequency | 1 times/15 minutes (can be customized) | 
| Upload Data | Product number + message type + product type number + data packet number + average leakage current within 15 minutes + peak leakage current within 15 minutes + leakage times within 15 minutes + remaining battery power + present detection date and time. | 
| Battery Life | 240 hous(When fully charged, set the interval for reporting data packets to 15 minutes) | 
| Power Supply | 3300AH | 
| Charge Mode | Integrated type, using charging seat to charge 5 current clamp at once time. | 
| Anti-theft function | Equip with built-in electronic lock, control the clamp lock or unlock by infrared remote controller | 
| LCD Size | 35mm×21.5mm;display area:32mm×15mm | 
| Indicator | Charging indicator: Infrared connector: Network indicator: alarm indicator | 
| Waterproof Level | IP65 of circuit part | 
| Insulation Strength | AC2KV/rms(between the iron core and case) | 
SELECTION GUIDE 1
Clamp on Leakage Current Monitoring Recorder
Click "Blue" marked MODEL NO.to view the spec
| Model | Range | Resolution | Clamp Size | Data Storge | Channels | Connection Type | 
| MEWOI8042 
 | 0.0mA~300.0A | 0.1mA | Ф40mm | - | - | Wireless | 
| MEWOI8072 
 | 0.0mA~1000A | 0.1mA | Ф70mm | - | - | Wireless | 
| MEWOI9000  | 0.0A-500A | 0.1A | 35mm×40mm | 4200Sets | Single | Wired | 
| MEWOI9000C  | 0.00mA-20A | 0.01mA | 25mm×30mm | 4200Sets | Single | Wired | 
| MEWOI9300  | 0.0A-500A | 0.1A | 35mm×40mm | 3900Sets | Three-Channels | Wired | 
| MEWOI9300C  | 0.00mA-20A | 0.01mA | 25mm×30mm | 3900Sets | Three-Channels | Wired | 
| MEWOI9500-130T  | 0.000mA-60.00A | 1uA | 25mm×30mm | 3000Sets | 1-8 Channels | Wireless | 
| MEWOI9500-140T  | 0.00mA ~ 300.0A | 0.01mA | 35mm×40mm | 3000Sets | 1-8 Channels | Wireless | 
| MEWOI9500-168T  | 0.00mA ~ 1200A | 0.01mA | φ68mm | 3000Sets | 1-8 Channels | Wireless | 
Technical Specification
| Functions |  Measuring AC leakage current and AC current, monitoring current in real-time manner, locating insulation faults and repairing  electrical lines. | 
| Power source |  DC6V, alkaline battery LR03 1.5V×4 | 
| Testing Method |  Clamp CT, TRMS | 
| Range |  0.00mA-20A | 
| Resolution |  0.01mA | 
| Accuracy |  ±2%±5dgt | 
| Clamp size | 25mm×30mm | 
| Instrument Dimension |  Tester(L×W×H): 75mm ×170mm ×30mm, Current clamp(L×W×H): 122mm ×70mm ×33mm | 
| Mass |  Tester: 410g (battery included),Current clamp: 170g | 
| LCD Dimension |  Display field: 44mm×27mm | 
| Shield Type |  Type B: Double shielding | 
| Sampling Rate |  2 samples/s | 
| Data storage |  3900 sets (data loss does not occur after power failure or replacing the battery) | 
| Time Setting |  Recording internal adjustable from 1 to 60 min, If set to 0 min, the auto-store function is turned off | 
| Recording Time |  Continuously operate for about 10 days at the power-saving mode | 
| Line Voltage |  Line with a voltage up to AC600V | 
| Overflow Display |  The “OL” symbol is displayed when a measured value is beyond the measurement range | 
| Battery. Voltage |  A Low-Battery symbol is displayed and the user should replace the battery with a new one when the battery voltage drops to 4.6V  to 4.8V. In this case, the measured values are still accurate.  | 
| Rated Power |  about 5mA at the power-saving mode, maximum current consumption: 20mA | 
| Lead Length |  2 m | 
| Temperature and Humidity |  Working: -20℃~50℃;below 80%rh | 
|  Storage: -20℃~50℃;below 70%rh | |
| Insulation Resistance |  Higher than 100MΩ, 1000V | 
| Safety Codes |  IEC1010-1, IEC1010-2-032, Pollution Class 2, CAT Ⅲ(600V), IEC61326(EMC standard) | 
| Accessories |  Tester: 1 PCS, Current clamp: 3PCS, Instrument bag: 1 PCS, RS232 Com-cable:1 PCS; Software Disk: 1 PCS | 
Tester: 1 PCS, Current clamp: 3PCS, Instrument bag: 1 PCS, RS232 Com-cable:1 PCS; Software Disk: 1 PCS
 
                                